We assess, free of charge and with no obligation, whether retrofitting your testing machine is feasible and worthwhile. Find out more
Free assessment of your machine Find out more
For test specimens according to EN 196-1 (40×40×160 mm cement-mortar prisms).
Patent pending, patent application No. 102023000026403 · European patent application No. 24218558.5
The failure pattern of the test specimen is an indicator of its maximum compressive strength: this is why it is important to ensure the best conditions for a uniform distribution of stress over the entire test surface of the specimen.
Devitech has developed an innovative sensor with 4 strain-gauge bridges that measures the force distribution over the area of the test specimen. Verifying the compression device (compression jig) is very difficult given the small size of the test area: the key element is checking the spherical seat.
The sensor is designed to measure how the load is distributed on the specimen. The signals of a load cell with the same dimensions as the cement-mortar specimen (40×40×40 mm), capable of measuring the loads acting on each side of the cube, are read and processed by the instrument, which displays and saves the acting loads.
The transducer is suitable for checking materials testing machines equipped with the EN 196-1 compression device.
By periodically monitoring how compressive stress is distributed over the specimen surface, better accuracy and repeatability of test results can be ensured.
The firmware shows, on the display of the instrument built into the carrying case, the 4 load values acting on the specimen surface and their average (AVG). Loads can be displayed in kN or MPa, along with the Rn% value of total load distribution across the 4 measuring bridges; acceptance limits can be set on three % levels (green, yellow, red) and values can be recorded to an external USB memory.
Lack of flatness and parallelism of the loading surfaces produces a non-axial thrust on the specimen.
Checking the efficiency of the compression platen’s spherical seat is essential to verify its ability to distribute the force uniformly over the specimen surface.
The Devitech sensor is placed in the machine in the same way and in the same position as the test specimen.
Instrumented carrying case
Sensor specifications
** maximum permissible values
This website uses cookies
We use technical cookies that the website needs to work and, only with your consent, statistics cookies (Google Analytics) and third-party content (Google Maps). You can accept, reject or choose which ones to enable, and change your mind at any time from the "Cookie preferences" link at the bottom of the page. Details in our cookie policy and privacy policy.